Spinal cord monitoring using evoked potentials recorded from feline vertebral bone
Autor: | Pablo Valencia, John C. Brown, Jens Axelgaard, Anders Nordwall, Donald R. Mcneal, Yoshio Harada |
---|---|
Rok vydání: | 1979 |
Předmět: |
Peripheral nerve stimulation
Reference Values medicine Animals Orthopedics and Sports Medicine Peripheral Nerves Evoked Potentials Electric stimulation Vertebral bone Spinal Cord Injuries Monitoring Physiologic business.industry Anatomy Spinal cord Spinal column Electric Stimulation Spine Functional integrity medicine.anatomical_structure Spinal Cord Reference values Cats sense organs Neurology (clinical) business Cancellous bone |
Zdroj: | Spine. 4(6) |
ISSN: | 0362-2436 |
Popis: | A new recording method for continuous spinal cord monitoring has been developed at Rancho Los Amigo Hospital. Using peripheral nerve stimulation, spinal evoked potentials (SEPs) are recorded from cancellous bone of spinous processes. This neurologically noninvasive technique provides a highly reproducible and consistent signal, unaffected by biologic noise and surgical handling of the spinal column. Moreover, the evoked response is extremely sensitive to changes in functional integrity of the spinal cord. In experimental spinal cord injuries, produced by localized distraction of the spine, changes of the evoked responses were well correlated with the change in spinal cord function. |
Databáze: | OpenAIRE |
Externí odkaz: |