9.4–14.1T small-animal PET-MR imaging: Feasibility analysis of LYSO APD readout via long signal lines

Autor: Joao Varela, Catarina Ortigao, R. Bugalho, Rolf Gruetter, Arthur W. Magill, J.A. Neves, Rui Silva, José C. Da Silva
Jazyk: angličtina
Rok vydání: 2013
Předmět:
Zdroj: Nuclear Instruments and Methods in Physics Research. A, Vol. 702 (2013) pp. 98-100
ISSN: 0168-9002
Popis: In the present work we intend to assess the readout feasibility of Avalanche Photodiode (APD) detectors via long signal lines for the development of a combined small-animal PET-MR prototype based on the ClearPEM technology. The detection performance of a LYSO-APD module was evaluated reading out the APD charge signals to the front-end ASIC via an 80 mm length flexible flat-cable (FFC). Experimental results showed a time resolution of 5.06 ns for the detector module in double-readout mode, with a nonsignificant degradation of 8.4% with the introduction of the FFC. The energy resolution of the system was not degradated by the FFC. (C) 2012 Elsevier B.V. All rights reserved.
Databáze: OpenAIRE