Occlusion tool software for pulmonary vein occlusion verification in atrial fibrillation cryoballoon ablation
Autor: | Filippo Maria Cauti, Gergana Shopova, Marco Polselli, Luigi Iaia, Pietro Rossi, Stefano Bianchi, Carlo Maria Giannitti, Alberto Arestia, Francesco Solimene, Giuseppe Stabile, Vincenzo Schillaci |
---|---|
Rok vydání: | 2020 |
Předmět: |
Epicardial Mapping
Male Cardiac Catheterization medicine.medical_treatment 030204 cardiovascular system & hematology Coronary Angiography Cryosurgery Sensitivity and Specificity Pulmonary vein 03 medical and health sciences 0302 clinical medicine occlusion tool Occlusion Image Processing Computer-Assisted medicine Humans Fluoroscopy atrial fibrillation Prospective Studies 030212 general & internal medicine pulmonary vein isolation medicine.diagnostic_test business.industry Cryoablation Atrial fibrillation General Medicine Middle Aged Ablation medicine.disease cryoablation new dielectric imaging system Contrast medium Catheter Pulmonary Veins Feasibility Studies Female Cardiology and Cardiovascular Medicine Nuclear medicine business Software |
Zdroj: | Pacing and Clinical Electrophysiology. 44:63-70 |
ISSN: | 1540-8159 0147-8389 |
Popis: | Background Optimal pulmonary vein (PV) occlusion, usually verified with selective contrast injection, is mandatory to obtain an effective PV isolation during cryoballoon (CB) ablation. Aim of the study The purpose of this study was to verify the feasibility and the accuracy of a new dielectric sensing system in assessing PV occlusion during CB ablation in patients with atrial fibrillation (AF). Methods We enrolled 28 consecutive patients with paroxysmal or persistent AF. After transseptal access, a detailed image reconstruction of left atrium and PVs was achieved with an octapolar or decapolar mapping catheter (Achieve catheter, Medtronic Inc.) and KODEX-EPD system (EPD Solutions, a Philips company). The degree of PV occlusion with the inflated Arctic Front Advance Cryoballoon (Medtronic Inc.) was verified using the new "occlusion tool" software module (EPD Solutions, a Philips company) and compared to an angiogram obtained with contrast medium injection in each PV. Results A total of 105 PV CB occlusion were tested. The new occlusion tool software module showed a 91% sensitivity and 76% specificity in assessing a complete PV occlusion verified with contrast medium injection. The positive predictive value was 80%, and the negative predictive value was 88.6%. Mean procedure time was 81 ± 17 minutes. Mean fluoroscopy time was 6 ± 2 minutes. No 30-day complications were observed. Conclusion The new dielectric imaging system was able to assess the degree of PV occlusion during a CB ablation with good sensitivity and specificity. |
Databáze: | OpenAIRE |
Externí odkaz: |