Computer assisted LISS plate placement: an in vitro study
Autor: | Jeffrey Yach, Andrew Quinn, Erin Janine Smith, David R. Pichora, Abdulaziz Al-Ahaideb, Randy E. Ellis |
---|---|
Rok vydání: | 2009 |
Předmět: |
Models
Anatomic Time Factors medicine.medical_treatment Bone Screws Condyle Fracture Fixation Internal Imaging Three-Dimensional Fracture fixation Bone plate medicine Fluoroscopy In vitro study Humans Femur Computer Simulation Mathematics Orthodontics Computer-assisted surgery medicine.diagnostic_test Anatomy Computer Science Applications Diaphysis medicine.anatomical_structure Surgery Computer-Assisted Surgery Family Practice Bone Plates |
Zdroj: | Computer aided surgery : official journal of the International Society for Computer Aided Surgery. 14(4-6) |
ISSN: | 1097-0150 |
Popis: | A laboratory study was conducted to compare the accuracy with which a LISS plate could be placed on the distal metaphysis of a model femur using both a fluoroscopy-based computer assisted technique and the conventional fluoroscopic technique. A significant difference was found between outcomes with the two approaches with respect to the maximum distance from the plate to the diaphysis of the bone, but there was no significant difference in the maximum distance to the condylar area. There was also no difference with respect to the number of holes that required re-drilling for adjustment of the plate placement or screws with poor purchase in bone. There were, however, significant differences between the two techniques in terms of duration of the procedure and radiation exposure. |
Databáze: | OpenAIRE |
Externí odkaz: |