Correction of error motion in a line-scanning tomographic optical microscope
Autor: | József Sinkó, Gábor Gajdátsy, Miklós Erdélyi, Laszlo Dudas, Gábor Szabó |
---|---|
Rok vydání: | 2012 |
Předmět: |
Physics
medicine.diagnostic_test business.industry Image quality ComputingMethodologies_IMAGEPROCESSINGANDCOMPUTERVISION Image processing Atomic and Molecular Physics and Optics Optics Optical coherence tomography Microscopy medicine Tomography Electrical and Electronic Engineering business Engineering (miscellaneous) Image resolution Rotation (mathematics) Instant centre of rotation |
Zdroj: | Applied optics. 51(26) |
ISSN: | 1539-4522 |
Popis: | A line-scanning tomographic optical microscope system requires precise rotation of the scanning line. Center of rotation error introduced by both the imprecision of optical and mechanical components is studied experimentally and via simulations. It was shown that a practical tolerance limit can be chosen where the influence of the investigated error on the reconstructed image quality remains insignificant. An effective and simply practical solution was presented to keep the center of rotation error below this tolerance limit and the spatial resolution of the reconstructed image close to the diffraction limit. |
Databáze: | OpenAIRE |
Externí odkaz: |