Prospective randomized comparative evaluation of proximal valve polyurethane and distal valve silicone peripherally inserted central catheters
Autor: | Shih C. Wang, Cheng K. Ong, Gabriel B. Lau, Sudhakar K. Venkatesh |
---|---|
Rok vydání: | 2008 |
Předmět: |
Adult
Male medicine.medical_specialty Catheterization Central Venous Time Factors Adolescent Radiography Polyurethanes Silicones Radiography Interventional Peripherally inserted central catheter Risk Assessment Comparative evaluation Upper Extremity chemistry.chemical_compound Young Adult Silicone Catheters Indwelling Risk Factors Catheterization Peripheral medicine Humans Radiology Nuclear Medicine and imaging Prospective Studies Vein Prospective cohort study Device Removal Aged Aged 80 and over Chi-Square Distribution business.industry Equipment Design Middle Aged Surgery Catheter medicine.anatomical_structure Logistic Models Treatment Outcome chemistry Catheter-Related Infections Catheter occlusion Equipment Failure Female Cardiology and Cardiovascular Medicine business Phlebitis |
Zdroj: | Journal of vascular and interventional radiology : JVIR. 21(8) |
ISSN: | 1535-7732 |
Popis: | The objective of this study was to evaluate and compare the relative durability and complications between the proximal valve polyurethane and distal valve silicone peripherally inserted central catheters (PICCs).Institutional review board approval was obtained. A total of 326 patients (mean age, 50.4 years) was assigned randomly to receive either a proximal valve polyurethane PICC (n = 198) or a distal valve silicone Groshong PICC (n = 194). All PICCs were inserted under radiologic guidance by interventional radiologists. Follow-up data were collected until catheter removal because of complications or treatment completion.The mean catheter dwell time was 25.6 days (range, 1-245 days). Complications were encountered in 26.8% and 47.9% of the proximal valve polyurethane PICCs and distal valve silicone PICCs, respectively (P.001). Significantly higher incidences of phlebitis (23.2% versus 11.6%, P = .003) and catheter-related infection (6.2% versus 2%, P = 0.043) were noted in the distal valve silicone PICCs. No significant differences in the incidence of catheter occlusion, fracture, or dislodgement were found. Multivariate logistic regression analysis showed a higher complication rate in the distal valve silicone PICCs corrected for patients' age, sex, underlying morbidity, indication, peripheral vein accessed, arm used, catheter tip placement, and the number of venepunctures attempted.Proximal valve polyurethane PICCs were more durable than distal valve silicone PICCs, which were associated with a higher incidence of phlebitis and infection, probably related to the materials of the catheters and the designs and placements of the catheter valves. |
Databáze: | OpenAIRE |
Externí odkaz: |