A case report of lead dysfunction presenting as high ventricular premature complex burden
Autor: | Ameya Udyavar, Saurabh Ajit Deshpande |
---|---|
Rok vydání: | 2022 |
Předmět: |
Lead dysfunction
Normal lead impedance medicine.medical_specialty Premature ventricular beats Lead revision biology business.industry Syncope (genus) Ventricular tachycardia medicine.disease biology.organism_classification RC666-701 Physiology (medical) Internal medicine Cardiology Diseases of the circulatory (Cardiovascular) system Medicine Cardiology and Cardiovascular Medicine business Lead (electronics) |
Zdroj: | Indian Pacing and Electrophysiology Journal, Vol 22, Iss 1, Pp 34-37 (2022) |
ISSN: | 0972-6292 |
DOI: | 10.1016/j.ipej.2021.10.003 |
Popis: | Pacemaker-troubleshooting is an important step in the evaluation of a patient with syncope post-pacemaker-implantation. The basic functions of sensing, pacing and impedance may remain spuriously normal in the case of lead-microfracture or insulation break. We report a case in which the lead dysfunction was diagnosed based on multiple episodes of premature ventricular beats. |