A semi-automatic planning technique for whole breast irradiation with tangential IMRT fields

Autor: L. Marrazzo, L. Redapi, M. Zani, S. Calusi, I. Meattini, C. Arilli, M. Casati, A. Compagnucci, C. Talamonti, D. Raspanti, S. Pertutti, V. Di Cataldo, L. Livi, S. Pallotta
Rok vydání: 2022
Předmět:
Zdroj: Physica Medica. 98:122-130
ISSN: 1120-1797
DOI: 10.1016/j.ejmp.2022.05.001
Popis: To implement a semi-automatic planning technique for whole breast irradiation with two tangential IMRT fields and to test the produced dose distribution against clinical 3DCRT plans, for introducing the technique in clinical practice.The Auto-Planning module of the PinnacleA statistically significant improvement in target coverage and dose homogeneity was observed for autoIMRT. No statistically significant differences were observed for ipsilateral organs, except for the ipsilateral lung in left DIBH, where slightly lower DThe developed autoIMRT technique was proven to be advantageous for target coverage and homogeneity and sufficiently robust towards isocenter displacements. The use of automated planning consistently reduces the planning workload with improvements in plan quality.