A practical workflow for capillary microsampling in nonclinical studies
Autor: | Raja Mangipudy, Ting Su, Linna Wang, Qin C Ji, Kristina D. Chadwick, Bonnie Wang, Renuka Pillutla |
---|---|
Rok vydání: | 2019 |
Předmět: | |
Zdroj: | Bioanalysis. 11:175-184 |
ISSN: | 1757-6199 1757-6180 |
DOI: | 10.4155/bio-2018-0224 |
Popis: | Aim: The result of investigation on the procedure of sample handling and bioanalysis of small volume of plasma sample for nonclinical studies stored in 0.5-ml micronic tubes was reported. Results/methodology: Sample integrity of the small volume (25 μl) during long-term storage and the feasibility and data reliability of performing multiple re-assays on the small volume sample using 5 μl aliquot per analysis was evaluated. Conclusion: Integrity was maintained in samples (25 μl) stored for up to 1 month in 0.5-ml micronic tubes at -20°C. A 25 μl sample is sufficient for four-times of re-assays. This evaluation demonstrated the feasibility of this workflow of handling and bioanalysis on small volume plasma sample for GLP studies under the US FDA guidance. |
Databáze: | OpenAIRE |
Externí odkaz: |