Analysis of photostimulable phosphor plate image artifacts in an oral and maxillofacial radiology department
Autor: | Wen-Chen Wang, Yuk-Kwan Chen, Shui-Hui Lin, Jin-Yi Chen, Hui-Lin Chiu, Chia-Hui Chen, Li-Min Lin |
---|---|
Rok vydání: | 2007 |
Předmět: |
Observer Variation
medicine.medical_specialty Photostimulated luminescence business.industry Computer science Radiography Radiography Dental Digital Oral and maxillofacial radiology Image (mathematics) Surgery Digital image Image Artifact Otorhinolaryngology medicine Humans Computer vision X-Ray Intensifying Screens Artificial intelligence Oral Surgery business Observer variation Artifacts General Dentistry |
Zdroj: | Oral surgery, oral medicine, oral pathology, oral radiology, and endodontics. 106(5) |
ISSN: | 1528-395X |
Popis: | Objective To analyze image artifact types and occurrence frequency when using a phosphor storage plate (PSP) digital radiographic system. Study design A total of 15,912 scanned digital images were evaluated by 3 observers, and image artifacts were classified into: 1) operator errors; 2) scanning errors; 3) PSP plate defects. To avoid damage to the sensor plate, a modification technique innovated by Roberts and Mol was used where needed and 2 double-sided pieces of tape were placed around the sensor plate covering. Results A total of 643 image artifacts were identified. The main image artifact cause was operator error (n = 554), followed by defects of plate (n = 60) and scanning (n = 29). Scanning errors could generally be corrected by rescanning, but most other artifacts required image retaking. The use of a modified clinical technique greatly reduced artifacts caused by wearing of the sensor plate (20%). Conclusions Modified methods to decrease the occurrence of image artifacts using PSP digital radiographic system are encouraged. |
Databáze: | OpenAIRE |
Externí odkaz: |