Evaluating the sensitivity of the optimization of acquisition geometry to the choice of reconstruction algorithm in digital breast tomosynthesis through a simulation study
Autor: | Kyle J. Myers, Predrag R. Bakic, Rongping Zeng, Subok Park |
---|---|
Rok vydání: | 2015 |
Předmět: |
Image quality
Computer science Normal Distribution Breast Neoplasms Geometry Iterative reconstruction Signal-To-Noise Ratio Imaging Three-Dimensional Image Processing Computer-Assisted Humans Computer Simulation Radiology Nuclear Medicine and imaging Computer vision Breast Sensitivity (control systems) Image resolution Observer Variation Likelihood Functions Artifact (error) Radiological and Ultrasound Technology Tomography X-Ray business.industry Reconstruction algorithm Digital Breast Tomosynthesis Simultaneous Algebraic Reconstruction Technique Female Artificial intelligence Noise (video) Artifacts business Algorithms Mammography |
Zdroj: | Physics in Medicine and Biology. 60:1259-1288 |
ISSN: | 1361-6560 0031-9155 |
DOI: | 10.1088/0031-9155/60/3/1259 |
Popis: | Due to the limited number of views and limited angular span in digital breast tomosynthesis (DBT), the acquisition geometry design is an important factor that affects the image quality. Therefore, intensive studies have been conducted regarding the optimization of the acquisition geometry. However, different reconstruction algorithms were used in most of the reported studies. Because each type of reconstruction algorithm can provide images with its own image resolution, noise properties and artifact appearance, it is unclear whether the optimal geometries concluded for the DBT system in one study can be generalized to the DBT systems with a reconstruction algorithm different to the one applied in that study. Hence, we investigated the effect of the reconstruction algorithm on the optimization of acquisition geometry parameters through carefully designed simulation studies. Our results show that using various reconstruction algorithms, including the filtered back-projection, the simultaneous algebraic reconstruction technique, the maximum-likelihood method and the total-variation regularized least-square method, gave similar performance trends for the acquisition parameters for detecting lesions. The consistency of system ranking indicates that the choice of the reconstruction algorithm may not be critical for DBT system geometry optimization. |
Databáze: | OpenAIRE |
Externí odkaz: |