Peritoneal dialysis catheters: repositioning with new fluoroscopic technique
Autor: | R L Siegel, L R Gesner, J L Nosher |
---|---|
Rok vydání: | 1994 |
Předmět: |
Adult
Male medicine.medical_specialty Adolescent medicine.medical_treatment Radiology Interventional Peritoneal dialysis Catheterization Catheters Indwelling Peritoneal Dialysis Continuous Ambulatory Female patient Medicine Humans Radiology Nuclear Medicine and imaging Dialysis Aged Aged 80 and over business.industry Middle Aged Surgery Catheter Equipment failure Fluoroscopy Equipment Failure Female Hemodialysis Catheter replacement business Pyelogram |
Zdroj: | Radiology. 190(3) |
ISSN: | 0033-8419 |
Popis: | From April 1990 to December 1992, the authors evaluated a new technique for radiologic manipulation of malfunctioning peritoneal dialysis catheters (PDCs). In 25 patients (16 male and nine female patients, aged 15-81 years), 25 PDCs were fluoroscopically manipulated 38 times for failure to drain (n = 22) or painful dialysis (n = 3). The PDC was backed out of its initial position with use of a wire and stiffener. Immediate success was obtained in 34 of 38 manipulations (89%), 21 manipulations (55%) were successful at 1 week, and durable success (function for more than 1 month or until the catheter was no longer needed) was achieved in 16 manipulations (42%). This is an effective alternative procedure to surgical catheter replacement or hemodialysis. |
Databáze: | OpenAIRE |
Externí odkaz: |