Physics::Medical Physics Proportional counter Radiation Dosage 030218 nuclear medicine & medical imaging Nuclear physics 03 medical and health sciences Radiation Protection 0302 clinical medicine Neutron generator Radiation Monitoring Humans Dosimetry Radiology Nuclear Medicine and imaging Neutron Graphite Neutrons Physics Range (particle radiation) Radiation Radiological and Ultrasound Technology Public Health Environmental and Occupational Health Gamma ray General Medicine Cesium Radioisotopes Gamma Rays 030220 oncology & carcinogenesis Neutron dosimetry |
Zdroj: | Radiation Protection Dosimetry. 180:168-171 |
ISSN: | 1742-3406 0144-8420 |
DOI: | 10.1093/rpd/ncx291 |
Popis: | The use of a custom-made cylindrical graphite proportional counter (Cy-GPC) along with a cylindrical tissue equivalent proportional counter (TEPC) for neutron-gamma mixed-field dosimetry has been studied in the following steps: first, the consistency of the gamma dose measurement between the Cy-TEPC and the Cy-GPC was investigated over a range of 20 keV (X-ray) to 0.661 MeV (Cs-137 gamma ray). Then, with both the counters used simultaneously, the neutron and gamma ray doses produced by a P385 Neutron Generator (Thermo Fisher Scientific) together with a Cs-137 gamma source were determined. |