A Comparison of Fresenius Com.Tec Cell and Spectra Optia Cell Separators for Autologous and Allogeneic Stem Cell Collections: Single Center Experience

Autor: Bülent Ündar, Şerife Solmaz, Selda Kahraman, Inci Alacacioglu, Celal Acar, Omur Gokmen Sevindik, Munire Turkyilmaz, Hayri Özsan, Mehmet Ali Özcan, Özden Pişkin, Fatih Demirkan
Jazyk: angličtina
Rok vydání: 2018
Předmět:
Popis: Peripheral blood is the prefered source for hematopoietic stem cells for hematopoietic stem cell transplantation. The efficiency of peripheral blood stem cell (PBSC) collection can vary among devices. In this study we aimed to compare feasibility and effectivity of apheresis procedures of the different systems. Two apheresis systems [Com.Tec (Fresenius Healthcare) and Spectra Optia (Caridian BCT)] were used in our center for the collection of PBSCs for autologous and allogeneic transplantation. We retrospectively analysed 190 apheresis procedures performed in healthy donors and patients between June 2012 and November 2014 in Department of Hematology, Dokuz Eylul University. PBSCS were collected by Fresenius cell separator (64 procedure) or Spectra Optia cell separator (126 procedure). Mobilization treatments were G-CSF (26.8%), cyclophosphamide plus G-CSF (48.4%), prelixafor plus G-CSF (14.7%), ESHAP (10%) and others. Patient and donor characteristics (age, weight, volume processed, disease, mobilization regimes) were similar in Fresenius and Spectra Optia apheresis groups. Altough both collected PBSCs efficiently, the amount of CD34+ cell in product collected by Spectra Optia device was significantly higher (p