Evaluation of an haematological analyser (Siemens Advia® 2120i) and improving quality of daily practice
Autor: | Magali Lefevre-Pettazzoni, Laura Melet, Christine Chaigneau |
---|---|
Rok vydání: | 2013 |
Předmět: |
Automation
Laboratory Risk Gynecology medicine.medical_specialty Hematologic Tests business.industry Cytodiagnosis Siemens Uncertainty Reproducibility of Results Professional Practice Professional practice General Medicine Quality Improvement Sensitivity and Specificity Accreditation Clinical Alarms Daily practice Equipment Contamination Humans Medicine business Retrospective Studies |
Zdroj: | Annales de biologie clinique. 71:679-691 |
ISSN: | 0003-3898 |
DOI: | 10.1684/abc.2013.0893 |
Popis: | Une analyse de risque, comprenant une verification de methode, ainsi qu’une etude de la pertinence des alarmes morphologiques declenchees par l’automate, ont ete realisees pour l’automate d’hematologie Advia ® 2120i (Siemens). La repetabilite, la fidelite intermediaire, la contamination, la linearite, la comparaison inter-automate, la justesse et l’approche de l’incertitude sont conformes aux resultats attendus (selon Ricos). Grâce a l’etude retrospective des alarmes morphologiques declenchees par l’automate et en s’inspirant des recommandations de l’ISLH, une nouvelle attitude concernant la relecture des frottis sanguins au microscope a ete instauree. Une sous-estimation du chiffre de plaquettes prelevees sur sang total citrate a egalement ete constatee malgre le facteur de correction de 1,1, par rapport a l’EDTA, necessitant probablement de redefinir des valeurs normales specifiques pour cet anticoagulant. |