A comparison between the use of a shadow shield whole body counter and an uncollimated gamma camera ain the assessment of the seven-day retention of SeHCAT

Autor: G. Phillips, C. L. Smith, B. R. Condon, G. Holdstock, P. J. Howlett, John S. Fleming, T. K. Hames, Duncan Ackery
Rok vydání: 1984
Předmět:
Zdroj: The British journal of radiology. 57(679)
ISSN: 0007-1285
Popis: We have compared the 7-day retention of the radioisotope bile salt analogue SeHCAT (75Se-23-selena-25-homotaurocholate), by whole body counting and by uncollimated gamma camera measurement, in phantoms and in 25 patients with inflammatory bowel disease. The results correlate with a linear correlation coefficient of 0.96. An uncollimated gamma camera can be used to assess bile acid malabsorption when a whole body radioactivity monitor is not available.
Databáze: OpenAIRE