Evaluation of Sensititre automated reading and incubation system for automated reading of Sensititre broth microdilution susceptibility plates
Autor: | Kimberle C. Chapin, Michael C. Musgnug |
---|---|
Rok vydání: | 2004 |
Předmět: | |
Zdroj: | Journal of clinical microbiology. 42(2) |
ISSN: | 0095-1137 |
Popis: | The Sensititre Automated Reading and Incubation System (ARIS) performs automated reading of Sensititre identification and susceptibility plates. ARIS susceptibility results were compared to manually read results. A total of 212 isolates (3,352 organism-antimicrobial agent combinations) yielded 99.4% essential agreement and categorical error rates of 0.2% minor, 0% major, and 1.0% very major. ARIS yields accurate susceptibility results compared to plates read manually. |
Databáze: | OpenAIRE |