A system for remote dosimetry audit of 3D-CRT, IMRT and VMAT based on lithium formate dosimetry
Autor: | Eva Lund, Sara Olsson, Åsa Carlsson Tedgren, Gudrun Alm Carlsson, Emelie Adolfsson, Håkan Gustafsson |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2014 |
Předmět: |
medicine.medical_specialty
Dosimeter Formates Radiotherapy business.industry Dosimetry audit remote audit end-to-end EPR ESR lithium formate Pilot Projects Radiotherapy Dosage Lithium formate Hematology Audit Imaging Three-Dimensional Oncology Audit system Medicine Dosimetry Radiology Nuclear Medicine and imaging Medical physics Radiotherapy Intensity-Modulated Radiologi och bildbehandling Radiometry business Nuclear medicine Radiology Nuclear Medicine and Medical Imaging |
Popis: | The aim of this work was to develop and test a remote end-to-end audit system using lithium formate EPR dosimeters. Four clinics were included in a pilot study, absorbed doses determined in the PTV agreed with TPS calculated doses within ±5% for 3D-CRT and ±7% (k=1) for IMRT/VMAT dose plans. |
Databáze: | OpenAIRE |
Externí odkaz: |