Application of time-frequency analysis to somatosensory evoked potential for intraoperative spinal cord monitoring
Autor: | K D K Luk, Jcy Leong, Yong Hu, William W. Lu |
---|---|
Rok vydání: | 2003 |
Předmět: |
Adult
Male Paper medicine.medical_specialty Adolescent Medical sciences Monitoring -Intraoperative-methods symbols.namesake Evoked-Potentials -Somatosensory-physiology Evoked Potentials Somatosensory Monitoring Intraoperative Reaction Time medicine Humans Latency (engineering) Child Analysis of Variance Fourier Analysis business.industry Short-time Fourier transform Reproducibility of Results Signal Processing Computer-Assisted Spinal cord Surgery Scoliosis surgery Time–frequency analysis Spinal-Cord-physiology Signal-Processing -Computer-Assisted Psychiatry and Mental health Amplitude medicine.anatomical_structure Scoliosis Spinal Cord Somatosensory evoked potential Fourier analysis Psychiatry and neurology medical sciences symbols Female Neurology (clinical) business Biomedical engineering |
Zdroj: | Journal of Neurology, Neurosurgery & Psychiatry. 74:82-87 |
ISSN: | 0022-3050 |
DOI: | 10.1136/jnnp.74.1.82 |
Popis: | Objective: To investigate the improvement in the reliability of intraoperative spinal cord monitoring by applying time-frequency analysis to somatosensory evoked potentials (SEP). Methods: 34 patients undergoing scoliosis surgery were studied. SEP were recorded during different stages of scoliosis surgery. Averaged SEP signals were analysed intraoperatively by short time Fourier transform (STFT). The time-frequency characteristics of SEP were observed during surgery. The main peak in the time-frequency interpretation of SEP was measured in peak time, peak frequency, and peak power. The changes in these variables were compared with the changes in latency and amplitude during different surgical stages. Results: During different surgical stages, changes in peak times and peak powers were found to correlate with the changes in latency and amplitude, respectively. Peak time showed more variability than latency (p < 0.01), while peak power showed less variability than amplitude (p < 0.01). The peak frequency of SEP appeared to be unchanged during surgery. SEP signals were found to have specific time-frequency characteristics, with the time-frequency distribution of the signals being located in a particular time-frequency space. Conclusions: Time-frequency analysis of SEP waveforms reveals stable and easily identifiable characteristics. Peak power is recommended as a more reliable monitoring parameter than amplitude, while peak time monitoring was not superior to latency measurement. Applying time-frequency analysis to SEP can improve the reliability of intraoperative spinal cord monitoring. published_or_final_version |
Databáze: | OpenAIRE |
Externí odkaz: |