Ex Vivo Performance of a Flexible Microwave Ablation Antenna

Autor: Susan C. Hagness, Nader Behdad, Yahya Mohtashami
Rok vydání: 2021
Předmět:
Zdroj: IEEE Transactions on Biomedical Engineering. 68:1680-1689
ISSN: 1558-2531
0018-9294
Popis: Objective: In this study, we investigate the performance of a flexible microwave ablation antenna for generating localized ablation zones. Methods: We designed a helical dipole antenna to operate at 1.9 GHz in egg white and liver. Semi-rigid prototypes of the antenna were fabricated and used to perform ablation experiments in egg white and perfused liver. Pulsed and continuous-wave power deliveries at different power levels were used. Flexible prototypes of the antenna were fabricated and used to perform ex vivo ablation experiments in perfused liver. Results: Pulsing was effective in reducing the shaft heating of semi-rigid cables. The antenna was capable of producing substantial ablation zones in perfused liver. Typical diameters (perpendicular to the antenna axis) of generated ablation zones with semi-rigid antennas in egg white and perfused liver were 30 mm and 20 mm, respectively. The flexible antenna had a good impedance match while bent. Average diameter of generated ablation zones by the flexible antenna with 10-W continuous-wave experiments in perfused liver was 26 mm. No significant difference was observed between the performances of semi-rigid and flexible prototypes. Conclusion: The flexible helical dipole antenna is capable of maintaining its good impedance match while bent and can generate substantial ablation zones in presence of perfusion. Significance: The proposed flexible antenna is promising for minimally invasive treatment of tumors that are otherwise inaccessible by rigid antennas. One example is lung where a catheter-based deployment of the flexible antenna into the tumor via airways may substantially reduce risks associated with using rigid antennas.
Databáze: OpenAIRE