Specific Absorption Rate and Specific Energy Dose: Comparison of 1.5-T versus 3.0-T Fetal MRI

Autor: N. Scott Adzick, Teresa Victoria, Hansel J. Otero, Christian A. Barrera, Camilo Jaimes, Michael S. Gee, Timothy P.L. Roberts, J. Christopher Edgar, Michael L. Francavilla, Suraj D. Serai
Rok vydání: 2020
Předmět:
Zdroj: Radiology. 295:664-674
ISSN: 1527-1315
0033-8419
DOI: 10.1148/radiol.2020191550
Popis: Background MRI performed at 3.0 T offers greater signal-to-noise ratio and better spatial resolution than does MRI performed at 1.5 T; however, for fetal MRI, there are concerns about the potential for greater radiofrequency energy administered to the fetus at 3.0-T MRI. Purpose To compare the specific absorption rate (SAR) and specific energy dose (SED) of fetal MRI at 1.5 and 3.0 T. Materials and Methods In this retrospective study, all fetal MRI examinations performed with 1.5- and 3.0-T scanners at one institution between July 2012 and October 2016 were evaluated. Two-dimensional (2D) and three-dimensional (3D) steady-state free precession (SSFP), single-shot fast spin-echo, 2D and 3D T1-weighted spoiled gradient-echo (SPGR), and echo-planar imaging sequences were performed. SAR, SED, accumulated SED, and acquisition time were retrieved from the Digital Imaging and Communications in Medicine header. Data are presented as mean ± standard deviation. Two one-sided tests with equivalence bounds of 0.5 (Cohen