Does experience influence the forces exerted on maxillary incisors during laryngoscopy? A manikin study using the Macintosh laryngoscope
Autor: | R. T M van Geel, C. Roberts, M. J. L. Bucx, J. T. Wegener, Theo Stijnen |
---|---|
Přispěvatelé: | Other departments |
Jazyk: | angličtina |
Rok vydání: | 1995 |
Předmět: |
Students
Medical Time Factors medicine.medical_treatment Laryngoscopy Dentistry Laryngoscopes Manikins Tongue Anesthesiology Maxilla Humans Medicine Intubation Nurse Anesthetists medicine.diagnostic_test business.industry Internship and Residency Equipment Design General Medicine Nurse anesthetist Incisor Anesthesiology and Pain Medicine medicine.anatomical_structure General Surgery Anesthesia Stress Mechanical business business.employer |
Zdroj: | Canadian journal of anaesthesia = Journal canadien d'anesthésie, 42(2), 144-149. Springer New York |
ISSN: | 0832-610X |
Popis: | The influence of the level of experience of the laryngoscopist on the duration of laryngoscopy, the forces exerted on the tongue and on the maxillary incisors during laryngoscopy, were investigated. Five groups (anaesthetists, residents in anaesthesia, nurse anaesthetists, surgeons and medical students), each consisting of 15 individuals, participated in the study. An intubation manikin was used with a laryngoscope modified so that the forces applied during laryngoscopy could be measured. The mean duration of laryngoscopy in these groups was 23.4 sec, 17.6 sec, 27.1 sec, 26.8 sec and 42.7 sec, respectively. The maximally applied forces on the tongue were 71.7 N, 60.5 N, 65.9 N, 74.2 N and 69.7 N, respectively. The maximally applied forces on the maxillary incisors were 49.9 N, 36.3 N, 41.1 N, 58.3 N and 53.9 N, respectively. These results indicate the level of experience has a significant influence on the duration of laryngoscopy but seems to have little influence on the forces applied to the tongue and the maxillary incisors. |
Databáze: | OpenAIRE |
Externí odkaz: |