Development and characterization of a scalable microperforated device capable of long-term zero order drug release

Autor: Zhiquan Luo, Phillip D. Bowman, Zhuojie Wu, Paul S. Ho, Salomon A Stavchansky, Ashish Rastogi
Rok vydání: 2010
Předmět:
Zdroj: Biomedical Microdevices. 12:915-921
ISSN: 1572-8781
1387-2176
DOI: 10.1007/s10544-010-9446-x
Popis: A drug delivery system that consists of microperforated polyimide microtubes was developed and characterized. Two groups of polyimide tubes were used. One set consisted of microtubes (I.D. = 125 microm) with 32.9 +/- 1.7 microm size holes. The second set consisted of larger tubes (I.D. = 1000 microm) with 362-542 microm holes. The number of holes was varied between 1 and 3. The small tubes were loaded with crystal violet (CV) and ethinyl estradiol (EE) and the drug release studies were performed in 0.01 M phosphate buffered saline (PBS) (pH 7.1-7.4) at 37.0 +/- 1.0 degrees C for upto 4 weeks. The large tubes were loaded with CV and the drug release was studied in vitro in PBS and also ex vivo in rabbit's vitreous humor. Linear release rates with R(2)0.9900 were obtained for all groups with CV and EE. Release rates of 7.8 +/- 2.5, 16.2 +/- 5.5, and 22.5 +/- 6.0 ng/day for CV and 30.1 +/- 5.8 ng/day for EE were obtained for small tubes. For large tubes, a release rate of 10.8 +/- 4.1, 15.8 +/- 4.8 and 22.1 +/- 6.7 microg/day was observed in vitro in PBS and a release rate of 5.8 +/- 1.8 microg/day was observed ex vivo in vitreous humor.
Databáze: OpenAIRE