Some technical factors affecting the intracranial pressure recorded by epidural space-occupying devices
Autor: | J. Zattoni, L. Semino, C. Siani |
---|---|
Rok vydání: | 1973 |
Předmět: |
medicine.medical_specialty
Intracranial Pressure medicine.diagnostic_test business.industry Transducers Interventional radiology Epidural device Epidural space Surgery medicine.anatomical_structure Medical Laboratory Science Methods Pressure medicine Humans Dura Mater Neurology (clinical) Neurosurgery Subdural space business Neuroradiology Intracranial pressure |
Zdroj: | Acta Neurochirurgica. 28:353-358 |
ISSN: | 0942-0940 0001-6268 |
DOI: | 10.1007/bf01405653 |
Popis: | Under controlled conditions, the possibility of recording from the epidural space a pressure higher than the pressure existing in the subdural space was reproduced. The phenomenon occurs when the epidural device records pressures deriving from a dural area larger than the area directly contacting the device. These results are obtained only when the epidural devices are shaped or placed in such a way to displace the surrounding dura from the bony face. |