On the use of pediatric phantoms in the dose evaluation during computed tomography (CT) thorax examinations
Autor: | F. Santopietro, G. Fariello, C. Furetta, V. Cannata, G. Giacco |
---|---|
Rok vydání: | 2001 |
Předmět: |
Thorax
medicine.medical_specialty medicine.diagnostic_test Phantoms Imaging business.industry Infant Newborn Computed tomography General Medicine Radiation Dosage Helical ct Imaging phantom Child Preschool Radiological weapon medicine Humans Dosimetry Radiography Thoracic Thermoluminescent Dosimetry Radiology Tomography Computed radiography Child Tomography X-Ray Computed business Nuclear medicine |
Zdroj: | Medical Physics. 28:199-204 |
ISSN: | 0094-2405 |
DOI: | 10.1118/1.1344205 |
Popis: | This paper presents the dosimetric results obtained during computed tomography (CT) thorax examinations, carried out using two different techniques, i.e., sequential CT (SQCT) and helical CT (HECT). The radiological examinations have been carried out on pediatric phantoms expressly designed and realized. Four different phantoms have been realized, according to the children shapes of 0, 2, 6, and 12 years old. The dosimetric measurements have been carried out using thermoluminescent dosimeters. The comparison of the dosimetric data obtained using the two different techniques shows that there is not any relevant difference between them as far as the entrance dose and the doses received by the critical organs are concerned. |
Databáze: | OpenAIRE |
Externí odkaz: |