Testing 144- and 256-crystal BGO block detectors
Autor: | C.W. Williams, Mark S. Andreaco, J.G. Rogers, R. Nutt |
---|---|
Rok vydání: | 1994 |
Předmět: |
Physics
Nuclear and High Energy Physics Physics::Instrumentation and Detectors business.industry Detector Gamma ray Particle detector Optics Nuclear Energy and Engineering Scintillation counter Calibration Measuring instrument High Energy Physics::Experiment Electrical and Electronic Engineering business Image resolution Block (data storage) |
Zdroj: | IEEE Transactions on Nuclear Science. 41:1423-1429 |
ISSN: | 1558-1578 0018-9499 |
DOI: | 10.1109/23.322926 |
Popis: | New block detectors have been fabricated incorporating large numbers of small crystals. The authors evaluate and compare the performance of the new detectors with a standard 64-crystal block detector from Siemens-CTI. The new detectors demonstrate greatly improved imaging capability for 511 keV gamma rays. Future PET tomographs incorporating such detectors should produce substantially better volume images with little increase in tomograph manufacturing costs. The detectors will require a new type of automatic calibration procedure. Various such procedures have been tested and are discussed. A technique using higher energy gamma rays has shown special promise. > |