Differential lung ventilation with a modified ventilator
Autor: | John J. Polanski, Deepak Vij, OʼNeal J. Sanders, William A. Conway, John Popovich |
---|---|
Rok vydání: | 1981 |
Předmět: |
Ventilators
Mechanical business.industry Peak Expiratory Flow Rate Forced Expiratory Flow Rates Critical Care and Intensive Care Medicine Intermittent Positive-Pressure Ventilation Intermittent positive pressure ventilation Bellows Deep breath Medicine Humans business Differential lung ventilation Tidal volume Biomedical engineering |
Zdroj: | Critical care medicine. 9(6) |
ISSN: | 0090-3493 |
Popis: | To achieve synchronized differential pulmonary ventilation, an Ohio 560 ventilator was modified through separation of the outputs of the deep breath bellows and the tidal volume bellows and providing independent volume, flow rate, PEEP, oxygen-humidification, and alarm systems. The described ventilator modification is simple, inexpensive, and does not alter the performance characteristics of the original unmodified ventilator. Differential lung ventilation may be more easily achieved with this modified ventilator than with the simultaneous use of two ventilators or external electronic synchronizers. |
Databáze: | OpenAIRE |