Popis: |
Use of metal-on-metal (MoM) articulations in total hip arthroplasty (THA) has sharply declined due to high failure rates from metal-related complications. Although certain MoM designs have demonstrated only 46% survival, not all MoM designs have performed the same. The purpose of this study is to evaluate mid-term to long-term survival of a specific MoM implant with a modular titanium taper adapter.A retrospective review was performed on all patients who underwent primary THA at our center with the MMean follow-up was 11.0 years (range 2-16; ±3.5). Mean cup angle of inclination was 42.8° (range 24°-70°, ±6.3°), with 88.0% reconstructed within the 40° ± 10° safe zone. There were 64 revisions (7.36%): 7 (0.81%) septic and 57 (6.56%) aseptic. Of those, 32 (3.68%) were adverse reactions to metal debris. Kaplan-Meier survival free of revision for all causes was 88.6% at 16 years (95% confidence interval 86.8-90.4). Univariate analysis of risk factors for all-cause, aseptic, and adverse reaction to metal debris revision found no relationship with female gender, age ≥65 years, body mass index30 kg/mThe results of this study demonstrate a more favorable mid-term to long-term survivorship with this specific MoM implant compared to other designs. Although our institution no longer performs MoM THA, further investigation into differences in MoM implant designs is warranted. |