Fiber-optic infrared radiometer for accurate temperature measurements
Autor: | Abraham Katzir, Ophir Eyal, Vered Scharf, Sharon Sade |
---|---|
Rok vydání: | 2002 |
Předmět: |
Radiometer
Optical fiber Materials science business.industry Materials Science (miscellaneous) Microwave radiometer Temperature measurement Industrial and Manufacturing Engineering law.invention Chopper Optics law Shutter Radiometry Business and International Management business Image resolution Remote sensing |
Zdroj: | Applied optics. 41(10) |
ISSN: | 1559-128X |
Popis: | A fiber-optic radiometer is developed for accurate noncontact temperature measurements. Of compact and novel design, it is based on replacing the usual chopper with a simple shutter. The radiometer operates in a spectral range of 5-20 microm and uses a silver-halide IR-transmitting optical fiber. The radiometer has a temperature resolution of 0.1 degrees C, a time response of 200 ms, and a spatial resolution of approximately 1 mm. Theory, simulation, radiometer design and construction, and examples of experimental measurements are shown. The novel radiometer can be used in diverse applications in science, medicine, and industry. |