Autor: |
A. Kyle Jones, Kevin A. Wunderle, Tom Fruscello, Michael Simanowith, Brendan Cline, Shalmali Dharmadhikari, Xinhui Duan, Jeremy C. Durack, David Hirschl, Don-Soo Kim, Usman Mahmood, Steve D. Mann, Charles Martin, Zeyad Metwalli, Jeffrey M. Moirano, Rebecca A. Neill, Janice Newsome, Horacio Padua, Alan H. Schoenfeld, Donald L. Miller |
Rok vydání: |
2023 |
Předmět: |
|
Zdroj: |
Journal of Vascular and Interventional Radiology. 34:556-562.e3 |
ISSN: |
1051-0443 |
Popis: |
To compare radiation dose index distributions for fluoroscopically guided interventions in interventional radiology from the American College of Radiology (ACR) Fluoroscopy Dose Index Registry (DIR-Fluoro) pilot to those from the Radiation Doses in Interventional Radiology (RAD-IR) study.Individual and grouped ACR Common identification numbers (procedure types) from the DIR-Fluoro pilot were matched to procedure types in the RAD-IR study. Fifteen comparisons were made. Distribution parameters, including the 10th, 25th, 50th, 75th, and 95th percentiles, were compared for fluoroscopy time (FT), cumulative air kerma (KThe median FT was equally likely to be higher or lower in the DIR-Fluoro pilot as it was in the RAD-IR study, whereas the maximum FT was almost twice as likely to be higher in the DIR-Fluoro pilot than it was in the RAD-IR study. The median KThe median dose indices have decreased since the RAD-IR study. The typical K |
Databáze: |
OpenAIRE |
Externí odkaz: |
|