Aortic valve
medicine.medical_specialty Transcatheter aortic medicine.medical_treatment Case Report AR - Aortic regurgitation Valve replacement Clinical Case medicine.artery Internal medicine Medicine AR aortic regurgitation Diseases of the circulatory (Cardiovascular) system Aorta TAVR transcatheter aortic valve replacement business.industry medicine.disease aortic valve Stenosis aorta medicine.anatomical_structure RC666-701 cardiovascular system Cardiology TAVR -transcatheter aortic valve replacement valve replacement Cardiology and Cardiovascular Medicine business |
Zdroj: | JACC: Case Reports, Vol 1, Iss 5, Pp 838-843 (2019) JACC Case Reports |
ISSN: | 2666-0849 |
Popis: | A 75-year-old woman with severe aortic stenosis underwent transcatheter aortic valve replacement using a 23-mm self-expanding Portico valve (Abbott). After complete deployment, it was found to be unacceptably low, with significant paravalvular regurgitation. The valve was pulled upward to the correct position successfully by using the snare maneuver. (Level of Difficulty: Advanced.) Graphical abstract A 75-year-old woman with severe aortic stenosis underwent transcatheter aortic valve replacement using a 23-mm self-expanding Portico valve… |