Reproducibility Materials science Biometry business.industry Ultrassonografia Ultrasound Axial length Catarata Standard deviation Cataract Ultrsonography Ophthalmology Optics Automotive Engineering Lenses intraocular Immersion (mathematics) Estudo comparativo Lentes intraoculares Surgery Intraocular lens power calculation Comparative study business |
Zdroj: | Revista Brasileira de Oftalmologia v.68 n.4 2009 Revista Brasileira de Oftalmologia Sociedade Brasileira de Oftalmologia (SBO) instacron:SBO Revista Brasileira de Oftalmologia, Volume: 68, Issue: 4, Pages: 212-215, Published: AUG 2009 |
Popis: | Objetivo: Comparar as medidas ecobiométricas pelos métodos de imersão e de contato. Métodos: A biometria ultrassônica foi realizada em 60 olhos, pelo mesmo examinador e com o mesmo aparelho, usando os métodos de contato e de imersão em cada olho. Resultados: A menor média do comprimento axial (23,16 mm) foi obtida através do método de contato, e a maior (23,29 mm) sob imersão, usando a mesma sonda ultrassônica. A média do desvio padrão entre as repetidas medidas no mesmo olho foi menor (0,02) com a técnica de imersão, e a maior (0,07) com o método de contato, diferença estatisticamente significante (p |