Multicenter trial in North America of UroLume urinary sphincter prosthesis
Autor: | W.C. Thorndyke, Jerzy B. Gajewski, Timothy B. Boone, Robert MacMillan, C. F Douglas Ackman, Caryn A. Abdill, David A. Rivas, Bruce Green, Joseph Binard, Anthony R. Stone, Albert Vazquez, Todd A. Linsenmeyer, Saad Juma, Rodney A. Appell, Michael P. Chetner, Michael B. Chancellor, Alfred Defalco, Janelle Foote, James K. Bennett, Claus G. Roehrborn, Michael E. Mayo |
---|---|
Rok vydání: | 1994 |
Předmět: |
Adult
Male medicine.medical_specialty Adolescent Urology medicine.medical_treatment Urination Prosthesis Dyssynergia Postoperative Complications Multicenter trial medicine Humans Urinary Bladder Neurogenic Spinal Cord Injuries Aged business.industry Urethral sphincter Stent Middle Aged Surgery Prosthesis Failure Urodynamics medicine.anatomical_structure Urethra Urinary Tract Infections Sphincter Urinary Sphincter Artificial business Detrusor sphincter dyssynergia |
Zdroj: | The Journal of urology. 152(3) |
ISSN: | 0022-5347 |
Popis: | We investigated the effectiveness and associated complications of treatment with an endoluminal urethral sphincter prosthesis in 153 spinal cord injury men (mean age 36 years, range 16 to 74 years) with urodynamically diagnosed detrusor-external sphincter dyssynergia. These patients were prospectively treated with a urethral sphincter stent at 15 centers in North America. Followup ranged from 2 to 33 months. Voiding pressures averaged 76 +/- 28, 42 +/- 21, 44 +/- 38, 35 +/- 18 and 32 +/- 20 cm. water, respectively, before prosthesis insertion in 153 patients and at 3 months in 123, 6 months in 114, 12 months in 98 and 24 months in 22. A significant decrease in voiding pressure was noted in the 22 patients at 24 months compared with matched preoperative data (80 +/- 25 cm. water, p = 0.03). The residual urine decreased from 181 +/- 154 ml. before insertion to 82 +/- 93 ml. at 24 months (p = 0.01). Maximum cystometric capacity remained constant, with a mean of 195 +/- 158 ml. before insertion to 248 +/- 122 ml. at 24 months (p = 0.17). No significant differences were apparent after 24 months of followup in any of the urodynamic parameters between 44 patients with and 109 without previous external sphincterotomy. Hemorrhage requiring blood transfusion, obstructive hyperplastic epithelial overgrowth and soft tissue erosion did not occur. No deleterious effects were observed on renal or erectile function. Of the patients 43 (28.1%) required 2 prostheses to bridge the external sphincter completely. Stent removal was required in 10 patients. Seven explantations were required for prosthesis migration, 1 for pain and urethral edema, 1 for inability to maintain a condom catheter, and 1 for nonepithelialization and secondary bladder neck obstruction. A total of 13 patients (8.5%) required a subsequent operation for bladder neck obstruction. Therefore, the sphincter prosthesis is an attractive modality for the treatment of external sphincter dyssynergia in patients with and without previous external sphincterotomy because of its ease of deployment and minimal associated morbidity. |
Databáze: | OpenAIRE |
Externí odkaz: |