Local Blood Flow in Human Leg Muscle Measured by a Transient Response Thermoelectric Method

Autor: Samuel A. Cucinell, William Perl
Jazyk: angličtina
Předmět:
Zdroj: Biophysical Journal. (2):211-230
ISSN: 0006-3495
DOI: 10.1016/S0006-3495(65)86712-1
Popis: The initial transient response of a Gibbs type thermoelectric probe embedded in human resting leg muscle was used for absolute quantitative measurement of local blood flow per unit tissue volume (local perfusion). The probe consisted of two thermistor-containing needles, one of which was heated by a constant electrical power input. The temperatures of both thermistors were recorded continuously on a two-channel, fast-response recorder. Upon sudden occlusion of the blood flow to the leg, each temperature vs. time record exhibited a change of slope. The change in slope of the temperature difference, divided by the temperature difference, (degrees/minute degree) was identified with the local perfusion (milliliters/minute milliliter) existing just before occlusion. The local perfusions determined agreed in range and mean with literature values of average perfusion by venous occlusion plethysmography. The nature of the local blood flow measured by the present method is discussed relative to that by other methods.
Databáze: OpenAIRE