An Assessment of the Corning Electrophoretic Procedure for High Density Lipoprotein Cholesterol Estimation
Autor: | T J N Carter, D G Bullock, Carmel M Cosgrove, S. Hughes |
---|---|
Rok vydání: | 1980 |
Předmět: |
Electrophoresis
Agar Gel 030213 general clinical medicine Chromatography Cholesterol Clinical Biochemistry 030209 endocrinology & metabolism General Medicine 03 medical and health sciences Electrophoresis chemistry.chemical_compound 0302 clinical medicine High-density lipoprotein chemistry Chemical Precipitation Humans Reagent Kits Diagnostic Ultracentrifuge Densitometer Lipoproteins HDL Ultracentrifugation Densitometry |
Zdroj: | Annals of Clinical Biochemistry: International Journal of Laboratory Medicine. 17:148-152 |
ISSN: | 1758-1001 0004-5632 |
DOI: | 10.1177/000456328001700309 |
Popis: | The electrophoretic procedure for high density lipoprotein cholesterol determination recently introduced by Corning Medical has been assessed. The results correlated well with those obtained by phosphotungstate/Mg precipitation and ultracentrifuge procedures, but the between-batch CV was 6% in the clinically important range below 1 mmol/l. The densitometer, which must be carefully aligned before use for this analysis, seemed sensitive to variations in the plates and patterns produced. The technique required great care, especially with timing, and appeared expensive. |