Autor: |
M. Lubin, R. Gerbelot, R. Prada, J. Porcherot, S. Bonnet |
Rok vydání: |
2021 |
Předmět: |
|
Zdroj: |
Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference. 2021 |
ISSN: |
2694-0604 |
Popis: |
This work presents a new dual-photoplethysmographic (PPG) system for pulse transit time (PTT) monitoring. An experiment has been set up in order to compare the PTT measurement between carotid and radial arteries from two systems: our physiological multimodal platform (PMP) and the Complior® tonometer. This work explores the comparison between such optical and mechanical modalities. The results show that the PPG device tends to overestimate the PTT (RMSE = 16 ms). Furthermore, both mechanical and optical signals have been superposed and demonstrated that pulse morphologies are quite similar.Clinical Relevance-Carotid-radial pulse wave velocity (PWV) is compared on a small cohort of subjects and significant differences are observed between optical and mechanical-based systems. |
Databáze: |
OpenAIRE |