Popis: |
Recently, biodegradable implants made from magnesium (Mg) alloys have been developed to obviate the need for later implant removal. Mg-based cannulated compression screws (CCS) are ideal for intramedullary screw (IMS) fixation of metacarpal fractures. The present study aimed at investigating the torque acting on Mg-based CCS at failure and at intramedullary metacarpal insertion. The devices were CE certified Magnezix 2.7 and 3.2 mm CCSs (Syntellix®, Hannover, Germany). Torque at failure was measured in a synthetic bone model using a standardized polyurethane foam block. In a second assessment, insertional torque was measured in ten cadaveric metacarpal bones. Mean torque at failure for the 2.7 mm and 3.2 mm CCSs was 42.8 Ncm (±1.9 Ncm) and 63.0 Ncm (±2.2 Ncm), respectively. In the human cadaver model, the torque distribution curve at metacarpal insertion showed three peaks. The highest reached 53.6% of the lowest torque at failure measured in the synthetic bone model for the 3.2 CCS (31.4 vs. 58.6 Ncm). The mean difference between peak torque at metacarpal insertion and torque at failure was 38.3 Ncm (99% CI [33.6, 43.0 Ncm], p 0.0001). In terms of torque load, Mg-based CCSs are suitable for IMS fixation of metacarpal fractures. Biodegradable implants may represent an important improvement of this treatment method; confirmation by in-vivo studies is needed. |