Evaluating the Precision of Preoperative Planning in Patient Specific Instrumentation: Can a Single MRI Yield Different Preoperative Plans?
Autor: | S. David Stulberg, Nitin Goyal |
---|---|
Rok vydání: | 2015 |
Předmět: |
Adult
Male Resection Software Image Processing Computer-Assisted Humans Medicine Computer Simulation Orthopedics and Sports Medicine In patient Femur Prospective Studies Instrumentation (computer programming) Femoral component Arthroplasty Replacement Knee Aged Aged 80 and over Preoperative planning Tibia business.industry Component sizing Reproducibility of Results Soft tissue Middle Aged Magnetic Resonance Imaging Treatment Outcome Surgery Computer-Assisted Preoperative Period Female Knee Prosthesis business Nuclear medicine Biomedical engineering |
Zdroj: | The Journal of Arthroplasty. 30:1250-1253 |
ISSN: | 0883-5403 |
DOI: | 10.1016/j.arth.2015.02.021 |
Popis: | If PSI preoperative planning were perfectly precise, a single MRI would give rise to one preoperative plan. Our purpose was to determine whether a single MRI inputted into two different PSI software yielded differences in preoperative alignment determination, component sizing, and bone resection within the preoperative plan. This prospective comparative study evaluated 40 preoperative plans generated by two PSI software given identical MRI. Femoral and tibial component sizes differed between software in 37.5% and 30.0% of cases, respectively. The maximum difference in bone resection between software ranged from 2.2mm to 5.1mm. Surgeons should be prepared to intraoperatively deviate from PSI selected size by 1 size. It may be necessary to fine tune soft tissue balancing when using a PSI system. |