Internally cooled antenna for microwave ablation: results in ex vivo and in vivo porcine livers
Autor: | Yongyan Gao, Yuanyuan Sun, Xiaoxia Ni, Lei Feng, Ping Liang, Yang Wang |
---|---|
Rok vydání: | 2007 |
Předmět: |
Parabolic antenna
Pathology medicine.medical_specialty Analysis of Variance business.industry Swine Microwave ablation General Medicine In Vitro Techniques Application time Liver In vivo Catheter Ablation Medicine Animals Radiology Nuclear Medicine and imaging Antenna (radio) business Microwaves Microwave Ex vivo Biomedical engineering |
Zdroj: | European journal of radiology. 67(2) |
ISSN: | 0720-048X |
Popis: | To evaluate the performance of microwave ablation with an internally cooled antenna in ex vivo and in vivo porcine livers.Microwave ablations were performed in ex vivo and in vivo porcine livers with a 1.9 mm-diameter 2450 MHz microwave needle antenna. Four power settings (50, 60, 70 and 80 W) were used in this study, while application time was fixed at 10 min. Diameters of the coagulation zone were observed on gross specimens.Excessive elongation of coagulation along the needle shaft was not encountered in all ablations. An arrow-shaped charring was observed. As the power increases, the long-axis coagulation diameter for in vivo liver increased significantly (P.05). The short-axis coagulation diameter for in vivo liver was significantly smaller than that for ex vivo liver (P.05) but not statistically different among the four power settings. After 10 min of ablation at 80 W, the short-axis and long-axis coagulation diameter for in vivo liver was 4.92+/-0.15 cm and 2.37+/-0.10 cm, respectively.The internally cooled microwave antenna may be advantageous to minimize collateral damages. |
Databáze: | OpenAIRE |
Externí odkaz: |