Common-path intrinsically achromatic optical diffraction tomography

Autor: Krzysztof Patorski, Maciej Trusiak, Malgorzata Kujawinska, Michał Ziemczonok, Michal Jozwik, Paweł Gocłowski, Julianna Winnik, Piotr Zdańkowski
Rok vydání: 2021
Předmět:
Zdroj: Biomed Opt Express
ISSN: 2156-7085
Popis: In this work we propose an open-top like common-path intrinsically achromatic optical diffraction tomography system. It operates as a total-shear interferometer and employs Ronchi-type amplitude diffraction grating, positioned in between the camera and the tube lens without an additional 4f system, generating three-beam interferograms with achromatic second harmonic. Such configuration makes the proposed system low cost, compact and immune to vibrations. We present the results of the measurements of 3D-printed cell phantom using laser diode (coherent) and superluminescent diode (partially coherent) light sources. Broadband light sources can be naturally employed without the need for any cumbersome compensation because of the intrinsic achromaticity of the interferometric recording (holograms generated by –1st and +1st conjugated diffraction orders are not affected by the illumination wavelength). The results show that the decreased coherence offers much reduced coherent noise and higher fidelity tomographic reconstruction especially when applied nonnegativity constraint regularization procedure.
Databáze: OpenAIRE