Popis: |
Purpose: The aim of this study was to investigate whether a chamber-type-specific radiation quality correction factor kQ can be determined in order to measure the reference air kerma rate of 60Co high-dose-rate (HDR) brachytherapy sources with acceptable uncertainty by means of a well-type ionization chamber calibrated for 192Ir HDR sources. Methods: The calibration coefficients of 35 well-type ionization chambers of two different chamber types for radiation fields of 60Co and 192Ir HDR brachytherapy sources were determined experimentally. A radiation quality correction factor kQ was determined as the ratio of the calibration coefficients for 60Co and 192Ir. The dependence on chamber-to-chamber variations, source-to-source variations, and source strength was investigated. Results: For the PTW Tx33004 (Nucletron source dosimetry system (SDS)) well-type chamber, the type-specific radiation quality correction factor kQ is 1.19. Note that this value is valid for chambers with the serial number, SN ≥ 315 (Nucletron SDS SN ≥ 548) onward only. For the Standard Imaging HDR 1000 Plus well-type chambers, the type-specific correction factor kQ is 1.05. Both kQ values are independent of the source strengths in the complete clinically relevant range. The relative expanded uncertainty (k = 2) of kQ is U kQ = 2.1% for both chamber types. Conclusions: The calibration coefficient of a well-type chamber for radiation fields of 60Co HDR brachytherapy sources can be calculated from a given calibration coefficient for 192 Ir radiation by using a chamber-type-specific radiation quality correction factor kQ . However, the uncertainty of a 60Co calibration coefficient calculated via kQ is at least twice as large as that for a direct calibration with a 60Co source. |