An Air-Operated Bistatic System for Breast Microwave Radar Imaging: Pre-Clinical Validation
Autor: | Tyson Reimer, Stephen Pistorius, Mario Solis-Nepote |
---|---|
Rok vydání: | 2019 |
Předmět: |
Computer science
Acoustics Breast Neoplasms 02 engineering and technology Iterative reconstruction Signal-To-Noise Ratio 030218 nuclear medicine & medical imaging law.invention 03 medical and health sciences 0302 clinical medicine Breast cancer law Radar imaging 0202 electrical engineering electronic engineering information engineering medicine Humans Breast Radar Microwaves Phantoms Imaging 020206 networking & telecommunications medicine.disease Bistatic radar Horn antenna Signal-to-noise ratio (imaging) Clutter Microwave |
Zdroj: | EMBC |
DOI: | 10.1109/embc.2019.8857646 |
Popis: | This study presents a breast microwave radar imaging system designed for bistatic operation in air. The system operates in the range of 1-8 GHz and employs a double-ridged horn antenna array with four degrees of freedom. A hemispherical breast phantom with a tumor inclusion of 1.5 cm diameter was used for validation. Reconstructed datasets resulted in artifact-free images where the tumor presence was detected. Signal to clutter ratios greater than 30 dB and tumor to fibroglandular ratios of 3 dB were measured. The encouraging images obtained with the system validate its potential as a clinical tool for breast cancer detection. |
Databáze: | OpenAIRE |