Acquisition Correction and Reconstruction for a Clinical SPECT/MRI Insert
Autor: | A. Savi, Botond Tolgyesi, Brian Hutton, Ilenia D'Adda, Rosa-Maria Moresco, A. J. Morahan, Kalman Nagy, Kjell Erlandsson, Carlo Fiorini, Debora Salvado, Marco Carminati, Z. Nyitrai, Pieter van Mullekom |
---|---|
Přispěvatelé: | Morahan, A, Erlandsson, K, D'Adda, I, Carminati, M, Savi, A, Moresco, R, Nagy, K, Nyitrai, Z, Tolgyesi, B, Salvado, D, Van Mullekom, P, Fiorini, C, Hutton, B |
Rok vydání: | 2019 |
Předmět: |
Scanner
medicine.diagnostic_test Pixel Computer science business.industry Detector Magnetic resonance imaging Iterative reconstruction Single-photon emission computed tomography Imaging phantom 030218 nuclear medicine & medical imaging 03 medical and health sciences 0302 clinical medicine 030220 oncology & carcinogenesis medicine Computer vision Artificial intelligence business Image resolution Clinical SPECT/MRI Insert |
Zdroj: | 2019 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC). |
DOI: | 10.1109/nss/mic42101.2019.9059982 |
Popis: | The development of the first clinical simultaneous Single Photon Emission Computed Tomography (SPECT) and Magnetic Resonance Imaging (MRI) system was carried out within the INSERT project. The INSERT scanner was constructed under the initial project, but its performance was not fully evaluated; here we have reconstructed the first images on the SPECT system. Calibration and acquisition protocols were developed and used to establish the clinical feasibility of the system. The image reconstruction procedures were implemented on the first phantom images in order to assess the system's imaging capabilities. This study solved issues involving incomplete data sets and pixel failure in the prototype detector system. The final images determined a measure of trans-axial image resolution, giving average values of 9.14 mm and 6.75 mm in the radial and tangential directions respectively. The work carried out on the complete system produced several clinical phantom images which utilized the capabilities of both SPECT and MRI. |
Databáze: | OpenAIRE |
Externí odkaz: |