Air-charged and microtransducer urodynamic catheters in the evaluation of urethral function
Autor: | Minda Neimark, G. Willy Davila, Jennifer T. Pollak, Jason T. Connor |
---|---|
Rok vydání: | 2003 |
Předmět: |
medicine.medical_specialty
business.industry Urology Concordance Transducers Balloon catheter Obstetrics and Gynecology Leak point pressure Middle Aged Urethral function Surgery Catheterization Catheter Urodynamics Urethral pressure profile Urethra medicine Pressure Humans Female Nuclear medicine business Maximum urethral closure pressure Aged |
Zdroj: | International urogynecology journal and pelvic floor dysfunction. 15(2) |
Popis: | This study aimed to compare measurements of urethral pressure profile and Valsalva leak point pressure (VLPP) obtained with air-charged and microtransducer catheters. Forty-five women with urogynecologic dysfunction underwent multichannel urodynamic evaluation including maximum urethral closure pressure (MUCP), functional urethral length (FUL), and VLPP with air-charged balloon catheters as well as microtransducer catheters. Lin’s concordance coefficient was used to examine the agreement of MUCP, VLPP, and FUL measurements with the two catheters. The MUCPs measured with the two catheters had a high concordance coefficient of 0.69 (95% CI 0.50, 0.82). The VLPP measurements obtained with the catheters also agreed well, with a concordance coefficient of 0.71 (95% CI 0.43, 0.87). The measurements of mean FUL had a low concordance of 0.35 (95% CI 0.085, 0.57). Overall, air-charged and microtransducer catheters yield similar information when evaluating VLPP and MUCP. There were differences in FUL and these were likely due to different catheter diameters. |
Databáze: | OpenAIRE |