A novel simplified method for managing inadvertent tube cut during aurolab aqueous drainage implant surgery for refractory glaucoma
Autor: | Sachin C Mungale, Paaraj Dave |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2019 |
Předmět: |
medicine.medical_specialty
Intraocular pressure Glaucoma medication genetic structures tube cut business.industry medicine.medical_treatment Glaucoma Case Reports medicine.disease eye diseases Surgery Ophthalmology Aurolab aqueous drainage implant Refractory Suture (anatomy) lcsh:Ophthalmology lcsh:RE1-994 medicine Tube (fluid conveyance) Implant business Ligature management |
Zdroj: | Indian Journal of Ophthalmology, Vol 67, Iss 5, Pp 694-696 (2019) Indian Journal of Ophthalmology |
ISSN: | 1998-3689 0301-4738 |
Popis: | We report a novel simplified method for managing inadvertent tube cut in a patient undergoing the Aurolab aqueous drainage implant (AADI) surgery for refractory neovascular glaucoma. Tube cut occurred while applying the polyglactin ligature suture used to avoid early postoperative hypotony. The short end of the cut tube was removed and the long end reinserted into the base plate of AADI. The surgery was then completed as usual. Following the surgery, the intraocular pressure reduced to 20 mmHg which stabilized and was maintained at 10 mmHg till the last follow-up at 12 months without any glaucoma medication. |