Evaluation of an infectivity standard for real-time quality control of human immunodeficiency virus type 1 quantitative micrococulture assays. Participating Laboratories of The AIDS Clinical Trials Group
Autor: | N Khan, D Brambilla, F. B. Hollinger, E Siwak, P S Reichelderfer, H Farzadegan, R W Coombs, C Beatty, S Rasheed, James W. Bremer, S M Hammer, Walter A. Scott, S A Fiscus |
---|---|
Rok vydání: | 1996 |
Předmět: | |
Zdroj: | Journal of Clinical Microbiology. 34:2312-2315 |
ISSN: | 1098-660X 0095-1137 |
DOI: | 10.1128/jcm.34.9.2312-2315.1996 |
Popis: | Quantitative microculture assays of cryopreserved human immunodeficiency virus type 1-infected cell suspensions and culture supernatants were compared among seven assays sites. There was no significant change in titer during 1 year of storage. The overall standard deviation for infected cell suspensions was approximately 0.8 log10 virus titer. A method for detecting deviant assay results was developed and was used to identify two donor cell preparations (n = 54) that gave consistently low titers. |
Databáze: | OpenAIRE |