Tribocorrosion: Ceramic and Oxidized Zirconium vs Cobalt-Chromium Heads in Total Hip Arthroplasty

Autor: James L. Howard, Sok Chuen Tan, Christopher Del Balso, Brent A. Lanting, Matthew G. Teeter, Adrian C.K. Lau
Rok vydání: 2016
Předmět:
Zdroj: Bone and Joint Institute
ISSN: 0883-5403
DOI: 10.1016/j.arth.2016.02.027
Popis: © 2016 Elsevier Inc. Background This matched-cohort study aims to compare tribocorrosion between matched ceramic and cobalt-chromium femoral head trunnions and between matched Oxinium and cobalt-chromium femoral head trunnions. Secondary objectives were to investigate whether taper design, depth of trunnion, implantation time, age, body mass index, and gender have an effect on fretting and corrosion. Methods All hip prostheses retrieved between 1999 and 2015 at one center were reviewed, giving a total of 52 ceramic heads. These were matched to a cobalt-chromium cohort according to taper design, head size, neck length, and implantation time. The trunnions were examined by 2 observers using a 4-point scoring technique and scored in 3 zones: apex, middle, and base. The observers were blinded to clinical and manufacturing data where possible. A separate matched-cohort analysis was performed between 8 Oxinium heads and 8 cobalt-chromium heads, which were similarly scored. Results Ceramic head trunnions demonstrated a lower median fretting and corrosion score at the base zone (P
Databáze: OpenAIRE