Fast calibration of electromagnetically tracked oblique-viewing rigid endoscopes
Autor: | Raj Shekhar, Xinyang Liu, Christina E. Rice |
---|---|
Rok vydání: | 2017 |
Předmět: |
Offset (computer science)
Computer science Camera matrix Biomedical Engineering Health Informatics Article 030218 nuclear medicine & medical imaging law.invention Telescope Magnetics 03 medical and health sciences 0302 clinical medicine law Camera auto-calibration Calibration Humans Radiology Nuclear Medicine and imaging Computer vision Image resolution Endoscopes Pixel business.industry Equipment Design General Medicine Computer Graphics and Computer-Aided Design Computer Science Applications Surgery Computer-Assisted 030220 oncology & carcinogenesis Surgery Computer Vision and Pattern Recognition Artificial intelligence business Electromagnetic Phenomena Algorithms Camera resectioning |
Zdroj: | Int J Comput Assist Radiol Surg |
ISSN: | 1861-6429 1861-6410 |
Popis: | The oblique-viewing (i.e., angled) rigid endoscope is a commonly used tool in conventional endoscopic surgeries. The relative rotation between its two moveable parts, the telescope and the camera head, creates a rotation offset between the actual and the projection of an object in the camera image. A calibration method tailored to compensate such offset is needed. We developed a fast calibration method for oblique-viewing rigid endoscopes suitable for clinical use. In contrast to prior approaches based on optical tracking, we used electromagnetic (EM) tracking as the external tracking hardware to improve compactness and practicality. Two EM sensors were mounted on the telescope and the camera head, respectively, with considerations to minimize EM tracking errors. Single-image calibration was incorporated into the method, and a sterilizable plate, laser-marked with the calibration pattern, was also developed. Furthermore, we proposed a general algorithm to estimate the rotation center in the camera image. Formulas for updating the camera matrix in terms of clockwise and counterclockwise rotations were also developed. The proposed calibration method was validated using a conventional $$30{^{\circ }}$$ , 5-mm laparoscope. Freehand calibrations were performed using the proposed method, and the calibration time averaged 2 min and 8 s. The calibration accuracy was evaluated in a simulated clinical setting with several surgical tools present in the magnetic field of EM tracking. The root-mean-square re-projection error averaged 4.9 pixel (range 2.4–8.5 pixel, with image resolution of $$1280 \times 720)$$ for rotation angles ranged from $$-40.3{^{\circ }}$$ to $$174.7{^{\circ }}$$ . We developed a method for fast and accurate calibration of oblique-viewing rigid endoscopes. The method was also designed to be performed in the operating room and will therefore support clinical translation of many emerging endoscopic computer-assisted surgical systems. |
Databáze: | OpenAIRE |
Externí odkaz: |