Determination of bromine and tin compounds in plastics using laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS)
Autor: | Murat Kayar, Belgin İzgi |
---|---|
Rok vydání: | 2014 |
Předmět: |
Detection limit
Bromine Chemistry Analytical chemistry chemistry.chemical_element Tin Compounds Mass spectrometry Mass Spectrometry Analytical Chemistry chemistry.chemical_compound Butyltin trichloride Bromide Calibration Tetrabromobisphenol A Laser Therapy Tin Inductively coupled plasma mass spectrometry Plastics Flame Retardants |
Zdroj: | Talanta. 139 |
ISSN: | 1873-3573 |
Popis: | The polybrominated flame retardants and organotin compounds were screened in terms of bromine and tin content using laser ablation inductively coupled plasma mass spectrometry in plastics. The calibration standards were prepared using the fused-disk technique, and all samples were investigated under optimal conditions. Using a central composite experimental design, laser parameters, laser energy, pulse rate, scan rate and spot size were identified. The detection limits of the method were 1000 mg kg(-1) and 1600 mg kg(-1) for bromide and tin, whereas the relative standard deviation (%) values of the analysis were 9% and 6% (n=3) for ERM EC681k with 770 +/- 70 mg kg(-1) Br and 86 +/- 6 mg kg(-1) Sn respectively, and 106-115% of Br and 102-104% of Sn were observed for the tetrabromobisphenol A and butyltin trichloride spike plastics, respectively. (C) 2015 Elsevier B.V. All rights reserved. |
Databáze: | OpenAIRE |
Externí odkaz: |