Commissural Alignment After Balloon-Expandable Transcatheter Aortic Valve Replacement Is Associated With Improved Hemodynamic Outcomes

Autor: Matthias, Raschpichler, Nir, Flint, Sung-Han, Yoon, Danon, Kaewkes, Chinar, Patel, Chetana, Singh, Vivek, Patel, Mohammad, Kashif, Michael A, Borger, Tarun, Chakravarty, Raj, Makkar
Rok vydání: 2022
Předmět:
Zdroj: JACC: Cardiovascular Interventions. 15:1126-1136
ISSN: 1936-8798
0231-8342
DOI: 10.1016/j.jcin.2022.04.006
Popis: Transcatheter aortic valve replacement (TAVR) is generally performed without control over the alignment of the bioprosthesis to the native aortic valve (AV) commissures. Data on the impact of commissural misalignment (CMA) on the clinical and hemodynamic outcome after TAVR are scarce.The aim of this study was to investigate the impact of commissural misalignment (CMA) on the clinical and hemodynamic outcome in patients with severe tricuspid aortic stenosis undergoing TAVR using the balloon-expandable (BE) SAPIEN 3 valve (Edwards LifeSciences).Clinical data of consecutive patients who underwent BE TAVR at Cedars-Sinai Medical Center (Los Angeles, California, USA) enrolled in the RESOLVE (Assessment of TRanscathetER and Surgical Aortic BiOprosthetic Valve Thrombosis and Its TrEatment With Anticoagulation) registry were retrospectively analyzed to evaluate CMA, which was defined as a neocommissure position30° compared with native commissures on computed tomography.A total of 324 patients (36.6% female, median Society of Thoracic Surgeons score of 3.9%) were included in the analysis. CMA was present in 171 individuals (52.8%). At 30 days, rates of aortic regurgitation greater than mild (5.6%) and a residual AV gradient ≥20 mm Hg (7.4%) were not different between CMA and non-CMA patients. Commissural orientation was independently associated with a relative AV mean gradient increase50% from discharge to 30 days (per increase of 10° misalignment; OR: 1.3; 95% CI: 1.0-1.4; P = 0.01). The long-term composite outcome of death or stroke was not different between groups (log-rank P = 0.29).In patients with severe tricuspid aortic stenosis who undergo SAPIEN 3 TAVR, the neocommissures align randomly. Our data demonstrate that commissural alignment may impact device performance and clinical outcomes in patients undergoing BE TAVR. (Assessment of TRanscathetEr and Surgical Aortic BiOprosthetic VaLVve Dysfunction with Multimodality Imaging and Its TrEatment with Anticoagulation [RESOLVE]; NCT02318342).
Databáze: OpenAIRE